Baxter SIGMA Spectrum Infusion Pumps: Multiple Vulnerabilities (FIXED)

Multiple vulnerabilities were found in Baxter SIGMA Spectrum infusion pumps and Wi-Fi battery units, including stored Wi-Fi credentials, format string flaws, and unauthenticated network reconfiguration. An attacker with physical or network access could retrieve Wi-Fi credentials, cause a denial of service, or redirect communications to a malicious gateway.

Deral Heiland Follow

Published on rapid7.com

Rapid7, Inc. (Rapid7) discovered vulnerabilities in two TCP/IP-enabled medical devices produced by Baxter Healthcare.

Rapid7, on rapid7.com

The full write-up lives on its author's site

It belongs to its author, so we link to it rather than copy it. The other two tabs are our own rewrites, checked against it.

Not written yet. The AI write-up for this write-up has not been written yet. The Original tab links to the source.

How it worked

Root cause

Likely a combination of storing sensitive credentials in non-volatile memory without encryption, using format string functions unsafely, and lacking authentication for network configuration commands.

Impact

An attacker could obtain Wi-Fi credentials, cause a denial of service, or man-in-the-middle communications, though some attacks require physical access or specific conditions.

What the attacker needed

Physical access to the device; network access to port 51243; ability to set up a rogue access point

The fix

Baxter has provided software updates to disable Telnet and FTP, address the format string vulnerability, and added authentication for network reconfiguration; organisations should also restrict physical and network access and properly decommission devices.

How it was found

  1. Obtain a Wi-Fi battery unit and connect it to the infusion pump.
  2. Power-cycle the infusion pump and remove the battery.
  3. Disassemble and reverse-engineer the battery to extract Wi-Fi credentials.
  4. For format string: set up a Wi-Fi access point with format string specifiers in the SSID.
  5. Send a get_wifi_location command via TCP or UDP port 51243 to the infusion pump.
  6. For unauthenticated reconfiguration: send an XML command 15 to TCP or UDP port 51243 to change the SIGMA GW IP address.

Explained by bugbounty.wiki with AI from the original and checked against quotes from it; it can still be wrong. Read the original before relying on a detail.

Not written yet. The AI write-up for this write-up has not been written yet. The Original tab links to the source.

These are security weaknesses in medical infusion pumps and their Wi-Fi batteries. Some require physical access to the device, while others can be triggered over the network. An attacker could steal Wi-Fi passwords, crash the device, or redirect its network traffic.

What is Info disclosure?

Data the application should keep private (personal data, internal details, other users' records) reaches someone it should not.

More by Deral Heiland

View all

1 more disclosed report, best first.

Similar reports

View all

Same bug class and techniques, clearest first.

Information disclosure

The lesson behind this report.

Data the application should keep private (personal data, internal details, other users' records) reaches someone it should not.