Skip to content
RCE Write-up Advanced

McAfee Enterprise ATR Uncovers Vulnerabilities in Globally Used B. Braun Infusion Pump

Five vulnerabilities in the B. Braun Infusomat Space infusion pump and SpaceStation, including an unauthenticated network protocol and a format string flaw, could allow an attacker to modify pump configuration and cause an unexpected dose of medication to be delivered to a patient.

mcafee.com

Published on mcafee.com

By Douglas McKee and Philippe Laulheret on Aug 24, 2021

The author, on mcafee.com

The full write-up lives on its author's site

It belongs to its author, so we link to it rather than copy it. The other two tabs are our own rewrites, checked against it.

Not written yet. The AI write-up for this write-up has not been written yet. The Original tab links to the source.

How it worked

Root cause

Likely the combination of an unauthenticated and unencrypted proprietary TCP protocol, an outdated third-party component with a known format string vulnerability, and missing authentication for critical functions allowed attackers to reach and modify pump configuration.

Impact

An attacker could modify the pump's configuration to deliver an unexpected dose of medication to a patient, potentially causing harm, with zero authentication required.

What the attacker needed

Network access to the pump's SpaceCom interface; pump in standby mode; knowledge of the proprietary protocol

The fix

B. Braun worked to adopt mitigations outlined in the disclosure report, but the text does not state that a fix was fully deployed.

How it was found

  1. Discover a remote interface at /rpc connected to json-dbus-bridge.
  2. Test for the known format string vulnerability in json-dbus-bridge.
  3. Exploit the format string vulnerability to gain www user shell access.
  4. Escalate privileges to root, possibly via a setuid binary or the backup/export feature.
  5. Use root access to interact with the CAN bus and send commands to the pump.
  6. Modify pump configuration, such as TUBE_HEADVOLUME, to cause an unexpected dose.

Explained by bugbounty.wiki with AI from the original and checked against quotes from it; it can still be wrong. Read the original before relying on a detail.

Not written yet. The AI write-up for this write-up has not been written yet. The Original tab links to the source.

An infusion pump is a medical device that delivers fluids and medicines into a patient's body at a controlled rate. The researchers found that the pump's communication module could be accessed over the network without any login, and that it ran outdated software with a known bug. By chaining these issues, an attacker could change how much medicine the pump would deliver, which could harm a patient.

What is RCE?

Running the attacker's code on the target's servers: through command or code injection, unsafe deserialisation or template injection.

Similar reports

View all

Same bug class and techniques, clearest first.

Remote code execution

The lesson behind this report.

Running the attacker's code on the target's servers: through command or code injection, unsafe deserialisation or template injection.